U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number M00146154B0
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
Event description: medical report: during the thrombectomy procedure with angiojet solent omni 6f, it was necessary to change the catheter because it stuck to the zipwire standard guide wire and it was not possible to remove it without exerting force and this caused damage to the catheter resulting in failure in use.After changing the procedure, it followed successfully.
 
Additional Manufacturer Narrative
The device was not received for evaluation at the time of this report; therefore, no physical analysis of the device could be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted in the device instructions for use (dfu) warning, "to prevent possible tissue damage, care should be taken when manipulating a device over a zipwire¿ during the device's placement and withdrawal.If resistance if felt during the device placement discontinue the procedure and determine the cause of the resistance before proceeding.If the cause of resistance cannot be determined, remove the zipwire hydrophilic guide wire and device as a unit to prevent possible damage and/or complications." additionally, the device instructions for use (dfu) preparation for use states, "note: if movement of the zipwire hydrophilic guide wire within the device becomes diminished, remove the guide wire and reactivate the coating by wetting its entire surface with heparinized saline solution." at this time, it is not possible to assign a definitive root cause for the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
VENUSA DE MEXICO S. DE R.L. DE C.V.
calle hertz 1525
parque industrial a j bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
victoria foss
340 lake hazeltine drive
chaska, MN 55318
9526418506
MDR Report Key15396238
Report Number9680001-2022-00111
Device Sequence Number6198695
Product Code DQX
UDI-Device Identifier08714729763918
UDI-Public08714729763918
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K000011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2024
Device Model NumberM00146154B0
Device Catalogue Number41537
Device Lot NumberJRZ6714956
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-