| Model Number |
1661G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.(b)(4).Related events captured via 2210968-2022-07404 and 2210968-2022-07406.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.During the 1st stitch, after passing the needle through the tissue and the suture was pulled, the suture broke at about half the position.A different lot of product was used to complete the procedure.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/23/2022.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned sample determined that a closed packet that pertains to product code 1661g, was received for evaluation.Upon initial inspection, of the sample, no external damages were observed on the packet.Visual inspection of the unopened sample, no defects were found on the packet.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed using an instron equipment and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related events captured via 2210968-2022-07404, 2210968-2022-07405 and 2210968-2022-07406.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional information: h4.Device manufacture date, h 6.Type of investigation.Additional information was requested, and the following was obtained: the original quantity involved for product code 1661g has been revised from (b)(4).The original quantity involved for product code 1832g has been revised from (b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|