| Model Number |
75023364 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a thr surgery, the polarcup trial insert nav 22/59 broke upon impaction.The procedure was completed, without delay, using a s+n back up device.Patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
D4: lot number updated.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a thr surgery, the polarcup trial insert nav 22/59 broke upon impaction.Pieces broke outside patient.The procedure was completed, without delay, using a s+n back up device.Patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: it was reported that, during a thr surgery, the polarcup trial insert nav 22/59 broke upon impaction.Pieces broke outside patient.Patient was not harmed.The complaint device intended for use in treatment and the broken-off piece was returned for investigation.A visual evaluation of the returned part was conducted, and it was concluded that one flap is broken-off and the device shows minor scratches on the surface.The device was effectively broken.So, the mentioned failure mode is confirmed.The performed complaint history review for the complaint device revealed no previous complaints for the same batch.A review of the production documentation did not show any deviations from the standard manufacturing procedure.A review of the risk management documentation verifies the failure mode and severity of the reported issue.Previous investigations demonstrated that the performance of the device is within the risks level that are anticipated in the risk management documentation.The current design remains therefore state of the art.As part of the continuous improvement, smith + nephew is nonetheless working on a design enhancement.No further escalation is required.Based on the performed investigations, the reported failure mode could be confirmed.A relationship between the reported event and the device can be confirmed.No probable cause can be determined.There is no indication that the reported devices failed to meet manufacturing specifications upon release for distribution.Normal wear and tear through repeated use is known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.This version of the device will be monitored for similar issues.This investigation is considered closed.The complaint device will be discarded.
|
| |
|