• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD VERSAPULSE POWERSUITE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD VERSAPULSE POWERSUITE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0638-803-00
Medical Device Problem Codes Defective Device (2588); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the console pedal sticks and continues to fire even when the surgeon removes his leg from the pedal.There is also a rattling sound inside the footswitch.There was no patient or procedure involved.
 
Additional Manufacturer Narrative
The console was not returned for analysis; however, this unit was serviced at the facility by a field service engineer (fse).The fse replaced the foot switch of the console system.Based on the work performed on the console, it is likely that the footswitch was damaged.Based on this information, the reported event is confirmed.The labelling review did not reveal any evidence of device off-label use or failure to follow instructions.The instruction for use (ifu) instructs that to prevent accidental laser discharge, always verify that the laser system is in standby mode before disconnecting the delivery system.Based on the information available, and console analysis results, an evaluation conclusion code of cause traced to component failure was assigned for the component damaged which could have affected the device performance and its integrity leading to the event experienced by the customer.
 
Event or Problem Description
It was reported that the console pedal sticks and continues to fire even when the surgeon removes his leg from the pedal.There is also a rattling sound inside the footswitch.There was no patient or procedure involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPULSE POWERSUITE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma st 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD
hakidma st 6
yokneam 20692 04
IS   2069204
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15396511
Report Number2124215-2022-35165
Device Sequence Number6198704
Product Code GEX
UDI-Device Identifier07290109145853
UDI-Public07290109145853
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0638-803-00
Device Catalogue Number0638-803-00
Device Lot Number002-47551
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer01/19/2023
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date02/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/06/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-