• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE¿ X  EP SYSTEM V1.1.1; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE¿ X  EP SYSTEM V1.1.1; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number ENSITE-SYSTEM-111
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the atrial fibrillation procedure, there was a display issue with the catheter resulting in cancellation of the procedure.It was believed to be a software issue.It was reported the ablation catheter appeared to miss-locate within the heart compared to other catheters.The tacticath se catheter was being used in navx mode.X-ray was attempted to continue the procedure, but it was ultimately decided to end the procedure and the patient was stable.The procedure has not been rescheduled.
 
Additional Manufacturer Narrative
Additional information: g3, h2, h3.The case study and collect log files were provided.The reported event stated "the ablation catheter appeared to mis-locate within the heart compared to other catheters" occurred.Review of the provided files indicate the sysref patch was placed quite low on the patient and pacing channels were on.Pacing channel distortion is larger when the catheter is farther away from the sysref patch and pacing channels are on.Ensure that the system reference patch is placed as close to the heart as possible and turn pacing channels off when not actively pacing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE¿ X  EP SYSTEM V1.1.1
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15396535
Report Number2184149-2022-00211
Device Sequence Number6124392
Product Code DQK
UDI-Device Identifier05415067041722
UDI-Public05415067041722
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberENSITE-SYSTEM-111
Device Lot Number8364128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/14/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age46 YR
Patient SexUnknown
-
-