Model Number IPN917532
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "after insertion, incomplete inflation of the balloon is noted on echocardiography".Additional information received stated that an additional catheter was used.Unknown if second catheter was inserted at the same site at time of report.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of iab would not inflate completely is not able to be confirmed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.Additional information received stated that the first insertion site was used for the second catheter.
 
Event or Problem Description
It was reported that "after insertion, incomplete inflation of the balloon is noted on echocardiography".Additional information received stated that an additional catheter was used.Unknown if second catheter was inserted at the same site at time of report.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBEROPTIX ULTRA 8 IAB: 8FR 40CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key15396774
Report Number3010532612-2022-00348
Device Sequence Number6198707
Product Code DSP
UDI-Device Identifier00801902136718
UDI-Public00801902136718
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K021462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/30/2023
Device Model NumberIPN917532
Device Catalogue NumberIAB-05840-LWS
Device Lot Number18F21M0033
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Patient SexUnknown
-
-