| Model Number |
21-3468-24 |
| Medical Device Problem Code |
Decrease in Suction (1146)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One (1) unit was received decontaminated without its original packaging.The unit was tested by introducing water with a syringe in order to verify for air in line.No air was detected and the water flowed without any problems.The complaint was not confirmed; thus, no root cause could be determined.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacturing.This remediation mdr was generated under protocol (b)(4), as a result of warning letter (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, when the nurse accessed the port for lab draw, the nurse was unable to aspirate blood and only a small amount was obtained with a syringe full of air.The patient was re-accessed with a new needle.No patient injury was reported.
|
| |
|