ST PAUL DELTEC GRIPPER MICRO BLUNT CANNULA, NON-CORING SA; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-2733-24
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One unit was received for evaluation; the returned unit was received decontaminated without its original packaging.The unit was tested for occlusion by introducing water with a syringe and found occlusion between the tube and luer.The failure mode reported is confirmed.As part of the standard process, production personnel perform a 100% visual inspection on the base for excess solvent and delamination.The root cause is excess solvent in the component bonding operation.Awareness notification was made to production personnel for the occlusion failure mode.Device history record (dhr) review relevant to the lot reported found no discrepancies or anomalies during manufacturing.This remediation mdr was generated under protocol (b)(4)., as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that the line would not prime past the top of the line/needle as if there is a blockage.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC GRIPPER MICRO BLUNT CANNULA, NON-CORING SA
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15397419
Report Number3012307300-2022-18452
Device Sequence Number5007310
Product Code FPA
UDI-Device Identifier10610586015847
UDI-Public10610586015847
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K870866
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2023
Device Model Number21-2733-24
Device Catalogue Number21-2733-24
Device Lot Number3716085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-