ST PAUL DELTEC GRIPPER MICRO BLUNT CANNULA, NON-CORING SA; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
21-2939-24 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One unit was received for evaluation; the returned unit was received decontaminated without its original packaging.The unit was tested for occlusion by introducing water with a syringe.The unit sample was found occluded between the tube and luer.The failure mode reported is confirmed.As part of the standard process, production personnel perform a 100% visual inspection on the base for excess solvent and delamination.The root cause is excess solvent in the component bonding operation.No lot number was provided; therefore, a device history record (dhr) review could not be completed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during the use of the product, the customer noticed the medical fluid was unable to be infused from the syringe to the needle; it could not be drawn.The customer checked the syringe and confirmed that the plunger moved without difficulty; therefore, it is suspected that there is an anomaly with the needle.No patient injury was reported.
|
| |
|