• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL DELTEC GRIPPER PLUS SAFETY NEEDLES; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL DELTEC GRIPPER PLUS SAFETY NEEDLES; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-2766-24
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned sample was received decontaminated; the sample was received in used condition with its original open package.The sample was tested with a syringe to replicate the failure mode; the complaint was not confirmed, no occlusion was detected.A device history record (dhr) review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr eview.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.
 
Event or Problem Description
It was reported that during the use of the product, the customer noticed medical fluid was unable to pass through it.Also, an occlusion was observed in the product.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC GRIPPER PLUS SAFETY NEEDLES
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
akasaka, minato-ku, tokyo
minneapolis, MN 55442
MDR Report Key15397427
Report Number3012307300-2022-18456
Device Sequence Number5007313
Product Code FPA
UDI-Device Identifier10610586025594
UDI-Public10610586025594
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/25/2025
Device Model Number21-2766-24
Device Catalogue Number21-2766-24
Device Lot Number4076583
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient RaceAsian
-
-