| Model Number |
21-2766-24 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during the use of the product, the customer noticed that the medical fluid was unable to pass through it; an occlusion was observed in the product.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One picture of the product was received.One unit was received decontaminated without its original packaging.The unit was tested for occlusion by introducing water with a syringe; no occlusion was detected and the failure mode reported was not confirmed.A device history record (dhr) review was not performed as no lot or serial numbers were provided.This remediation mdr was generated under protocol(b)(4), as a result of warning letter cms#(b)(4).
|
| |
|