• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL DELTEC VALVED PEEL AWAY SHEATH INTRODUCER; PORT CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL DELTEC VALVED PEEL AWAY SHEATH INTRODUCER; PORT CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCU Back to Search Results
Model Number 21-4477-24
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Samples were returned for evaluation: one sample was visually inspected at a distance of 12 in to 16 in under normal conditions of illumination; no discrepancies were detected; the other sample was received with a catheter piece trapped in the sleeve.Functional test was performed on the sample without the catheter trapped according to instructions for use (ifu): the catheter was connected to the port in order to reproduce the failure mode or find an unusual function; then it was pulled in order to verify if the catheter would detach easily.No discrepancies were detected, the catheter remained attached after pulling it.Root cause could not be determined since the complaint was not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacturing.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that it was not possible to attach the catheter with the connector to the port housing coupling device.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC VALVED PEEL AWAY SHEATH INTRODUCER
Common Device Name
PORT CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCU
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15397524
Report Number3012307300-2022-18480
Device Sequence Number5057236
Product Code LJT
UDI-Device Identifier10610586039997
UDI-Public10610586039997
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K072657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2024
Device Model Number21-4477-24
Device Catalogue Number21-4477-24
Device Lot Number3872422
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Initial Report FDA Received Date09/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-