| Model Number |
21-2762-24 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the needle was not continuous and completely locked.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One unit was received decontaminated without its original packaging.It was tested for occlusion by introducing water with a syringe.The unit sample was found occluded; the failure mode reported is confirmed.The root cause is excess solvent in the component bonding operation.Awareness notification was made to production personnel for the occlusion failure mode.A device history record (dhr) review indicated no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|