• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CAPNOCHECK II, MODEL 8400; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CAPNOCHECK II, MODEL 8400; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 8400
Medical Device Problem Codes Device Alarm System (1012); Failure to Power Up (1476); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product sample was received for evaluation.Visual and functional testing were performed.Upon visual inspection damage to the physical casing could be observed.The reported problem was able to be duplicated through testing the device.The root cause of the reported issue was found to be a defective main board and bench.The damaged co2 pump and broken case is consistent with unit being severely impacted from drop.Both the main board and the bench were replaced.Manufacturing device history record (dhr) review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.Root cause user interface from complications from the device being dropped.Operator of device was unknown.No information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that there was a rattling inside the device.It was also reported the device exhibited memory check error with a small black spot on the screen, alarm bell, and the yellow and red lights would flash.It was reported that the only way to stop this alarm was to remove the battery.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPNOCHECK II, MODEL 8400
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15397570
Report Number3012307300-2022-18498
Device Sequence Number6102639
Product Code CCK
UDI-Device Identifier10610586036354
UDI-Public10610586036354
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8400
Device Catalogue Number8400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Initial Report FDA Received Date09/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-