| Model Number |
21-2966-24 |
| Medical Device Problem Code |
Reflux within Device (1522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned samples (10) were received decontaminated with its original closed packaging.The samples were tested with a syringe to replicate the failure mode.No occlusion was detected; the complaint was not confirmed and root cause could not be determined.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|
| |
|
Event or Problem Description
|
|
It was reported that there was no blood reflux despite attempts with the same sample.There were no issues identified the previous times.No patient injury was reported.
|
| |
|