| Model Number |
10-390 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Email - (b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: h3, h6 device evaluation: the device related to the reported event of rupture was received on (b)(6) 2022 with lot number 2666408.Based on the device analysis grid, the assessments of the complaint are: rupture: was observed, broken device on the radius side assessed as surgical impact opening (shell thickness within specification).Additional observations: creases are observed.Stress marks are observed assessed as non-penetrating nick.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left-side rupture.Device has been removed and replaced.
|
| |
|