• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION®; DIMENSION® (LTNI) CARDIAC TROPONIN I FLEX® REAGENT CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION®; DIMENSION® (LTNI) CARDIAC TROPONIN I FLEX® REAGENT CARTRIDGE Back to Search Results
Model Number 10444896
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A united states customer contacted the siemens healthcare diagnostics remote services center (rsc) regarding a discordant falsely elevated cardiac troponin-i (ltni) patient sample result obtained on a dimension xpand plus with hm system.Siemens is investigating the event.
 
Event or Problem Description
A discordant falsely elevated cardiac troponin i (ltni) result was obtained on a patient sample on a dimension xpand plus with hm system.The result was reported to the physician.The patient was sent to an alternate facility where a new sample was drawn and processed on an alternate siemens system.A lower result, considered correct, was obtained and reported.There are no known reports of adverse health consequences due to the discordant falsely elevated ltni patient result or the transfer to an alternate facility.
 
Additional Manufacturer Narrative
Additional information (14-sep-2022): siemens healthcare diagnostics headquarters support center (hsc) concluded the investigation of the event.The customer reported that a falsely elevated result was obtained on 24-aug-2022 with the dimension cardiac troponin i (ltni) assay lot ea2314 on a lithium heparin plasma patient sample.Siemens healthcare diagnostics inc.Has investigated discordant falsely elevated results with cardiac troponin i (ltni) on the dimension xpand plus system with lot ea2314.Siemens has confirmed that dimension cardiac troponin i (ltni) lot ea2314 shows a positive bias with lithium heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/ml [0.07 ug/l].Troponin values as high as 0.20 ng/ml [0.20 ug/l] have been observed with lithium heparin plasma that are expected to be within the 99th percentile.Serum samples collected from normal, healthy individuals whose results fall within the reference interval or below the 99th percentile of 0.07 ng/ml [0.07 ug/l] were not impacted by this issue.Urgent medical device correction, dc-22-02.A.Us dated may 18, 2022 and urgent field safety notice dc-22-0-02.A.Ous dated may-2022 was sent to all customers who had been shipped dimension cardiac troponin i (ltni) lot ea2314.Siemens recommended that customers use serum patient samples only when processing samples with ltni lot ea2314.The customer was instructed to discard ltni lot ea2314 if their laboratory is not able to use serum samples.Headquarters support center (hsc) confirmed that the customer acknowledged they had received the may 2022 urgent medical device correction (umdc) letter prior to the event.The customer was instructed in the umdc to not use plasma samples when using the ltni assay lot ea2314 or to process samples with an alternate lot if plasma had to be used as a sample type.Operator error contributed to the event.Section h6 has been updated to reflect siemens' investigation.Initial mdr was filed on 07-sep-2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION®
Common Device Name
DIMENSION® (LTNI) CARDIAC TROPONIN I FLEX® REAGENT CARTRIDGE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer Contact
james morgera
500 gbc drive
po box 6101
newark, DE 19714-6101
MDR Report Key15401068
Report Number2517506-2022-00238
Device Sequence Number5057338
Product Code MMI
UDI-Device Identifier00842768013904
UDI-Public00842768013904
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K010313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2022
Device Model Number10444896
Device Catalogue NumberRF 521 SMN 10444896
Device Lot NumberEA2314
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2517506-05/24/2022-002-C
Patient Sequence Number1
Patient SexUnknown
-
-