MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH 5.5 TI CORT FIX 8X40MM; ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH 5.5 TI CORT FIX 8X40MM; ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION Back to Search Results
Model Number 186731840
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional product codes: kwq, mni,kwp and nkb.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Initial reporter name and address: (b)(6).Reporter is a j&j sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that this was a plif procedures (l4/5) for treating asd on (b)(6) 2022.The event took place in sequence as follows: there was a possibility that the screw 7.0×40 (unk), that inserted to l5 in this surgery was loose, so it was removed.The surgeon tried to replace it with the screw 8.0×40 and tried to insert it but was unable to insert it due to resistance.The tip of the screwdriver in question was broken.The tip of the screwdriver got caught in the core of the screw 8.0×40, and the surgeon had difficulty removing the screw 8.0×40, but he managed to do so and the removal was completed.The surgeon commented that the reason for the difficulty in inserting the screw 8.0×40 was that the patient's bone was quite hard, and that when he did not tap the screw 8.0×40 to break through although he felt resistance during insertion.The surgery was completed successfully within 30 minutes delay.No further information is available.This report is for one (1) 5.5 ti cort fix 8x40mm.This is report 2 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: a manufacturing record evaluation was performed for the finished device.Product code: 186731840.Lot number: 256315.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 04/10/2019.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the 5.5 ti cort fix 8x40mm, p/n: 186731840, exhibits signs of normal use.No significant product problem was observed on the surface of the complaint device.A dimensional inspection was unable to be performed due to the complex geometry of the device.A functional test was not conducted due to the mating device being returned in a broken condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the 5.5 ti cort fix 8x40mm, p/n: 186731840 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 TI CORT FIX 8X40MM
Common Device Name
ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 02400
SZ   02400
3035526892
MDR Report Key15401501
Report Number1526439-2022-01566
Device Sequence Number5183747
Product Code MNH
UDI-Device Identifier10705034352183
UDI-Public(01)10705034352183
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110216
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186731840
Device Catalogue Number186731840
Device Lot Number256315
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 TI CORT FIX 7X40MM; VIPER UNIVERSAL POLY DRIVER
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-