MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FEM LT SZ 5 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FEM LT SZ 5 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 1504-00-125
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that when preparing for implantation, upon opening the sterile packaging of the femur.There was a foreign object on the back of the implant.We immediately opened a new implant for the surgery.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that there was no any adverse consequences that affected the patient because of the reported event.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The information received, will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed.And the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed, for the finished device 150400125, lot#: 9823806.And no non-conformances/manufacturing irregularities were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned for analysis.Visual analysis revealed that the device has signs of use, the device was returned without of the inside sterile packaging.Due to its not the first use of the device and no more evidence were received the report allegation cannot be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device 150400125, lot 9823806 and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM LT SZ 5 NAR CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15401814
Report Number1818910-2022-17653
Device Sequence Number5098303
Product Code NJL
UDI-Device Identifier10603295041146
UDI-Public10603295041146
Combination Product (Y/N)N
Initial Reporter CountrySN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-00-125
Device Catalogue Number150400125
Device Lot Number9823806
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer09/27/2022
10/18/2022
01/18/2023
02/03/2023
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date09/30/2022
10/19/2022
01/20/2023
02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-