Model Number A701157
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
Related manufacturing ref: 3008452825-2022-00596.During a premature ventricular contraction procedure (pvc), two ablation catheters showed abnormal impedance and signal interference which caused a delay.While using the tacticath¿ contact force ablation catheter, sensor enabled to discharge the left ventricular outflow tract for 30 seconds, impedance increased from 105 to over 150, accompanied by signal interference and the radiofrequency instrument automatically cut off the discharge.A flexability¿ablation catheter, uni-d was then used which started normally, however, the same issue occurred approximately 15 minutes later.The catheter was replaced again and the procedure was completed with no consequences to the patient.
 
Additional Manufacturer Narrative
One quadripolar, uni-directional, curve d, flexability ablation catheter was received for evaluation.Electrode 1 met specifications for acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported impedance issue and subsequent delay remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXABILITY¿ABLATION CATHETER, UNI-D, CURVE D
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15402024
Report Number3008452825-2022-00597
Device Sequence Number20085384
Product Code OAD
UDI-Device Identifier05415067011497
UDI-Public05415067011497
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberA701157
Device Catalogue NumberA701157
Device Lot Number8296943
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SE
Patient SexUnknown
-
-