| Model Number |
MI1000 MED-EL CONCERT PIN |
| Medical Device Problem Codes |
Impedance Problem (2950); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user suffered from a head trauma while playing at school and has no hearing sensation since.Re-implantation is considered, but not scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to the reported trauma appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user suffered from a head trauma while playing at school and has no hearing sensation since.Re-implantation is considered, but not scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user_s hearing performance with the device is affected.The user suffered from a head trauma while playing at school and has no hearing sensation since.The user has been re-implanted.
|
| |
|