(b)(4).Performed a visual inspection of the returned item and found it to exhibit signs of repeated use nicked / gouged and has fractured on the lateral side.The device history records & receiving inspection report were reviewed for deviations and/ or anomalies with no deviations/ anomalies identified.Medical records were not provided.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.In addition to the visual and optical microscopy analysis, a review of the dhf and dfmea was performed.Ongoing complaint monitoring confirms that the rate of instrument fracture is within the expected risk profile as specified in the dfmea.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|