• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number LS1500
Medical Device Problem Codes Peeled/Delaminated (1454); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, the white rubber material (insulation) attached to the tip of the jaws (narrow part between the tip and the hinge) were partially not attached and peeling off but, it did not fall into the patient's abdominal cavity.However, the broken piece was not found.A new device was used and worked as usual.
 
Additional Manufacturer Narrative
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a damaged/missing insulation.It was reported that the device had a component that disengaged and the insulation of the device was damaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: carefully insert and withdraw instruments from cannulas to avoid possible damage to the devices and/or injury to the patient.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic cholecystectomy procedure, the white rubber material (insulation) attached to the tip of the jaws (narrow part between the tip and the hinge) were partially not attached and peeling off but, it did not fall into the patient's abdominal cavity.However, the broken piece was not found.A new device was used and worked as usual.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CH  201114
Manufacturer (Section G)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CH   201114
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15402579
Report Number3006451981-2022-00087
Device Sequence Number5183279
Product Code GEI
UDI-Device Identifier30884524000279
UDI-Public30884524000279
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K031011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2026
Device Model NumberLS1500
Device Catalogue NumberLS1500
Device Lot NumberS1FB001X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer09/13/2022
09/13/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date09/13/2022
09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-