| Model Number |
MAJ-1444 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report was not returned to olympus for evaluation.The definitive cause of the user's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
|
| |
|
Event or Problem Description
|
|
The customer reports during a laparoscopic ultrasound endoscopy procedure using an evis lucera ultrasound gastrovideoscope with an air/water valve for the treatment of bleeding there was no air supply at all.The procedure was aborted because there was no improvement after withdrawing the scope and checking it.Hemostasis was achieved that day using another scope.The next day, the physician still needed the ultrasound function, so it was decided to use the scope again.Although there were no abnormalities in the pre-use check, air supply became impossible during the procedure, and it was decided to cancel it.From the next day onwards, after new equipment arrived, the procedure was carried out and completed without any problem.There is no impact or health hazard to the patient related to stopping the procedure due to the equipment issue.Case with patient identifier (b)(6) reports the scope used in the first procedure.Case with patient identifier (b)(6) reports the air/water valve in the first procedure.Case with patient identifier (b)(6) reports the scope used in the procedure the following day.Case with patient identifier (b)(6) reports the air/water valve used in the procedure the following day.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The dhr was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the air and water supply button was tested in combination with the scope of the indicated phenomenon, but we were able to confirm damage to the device.As a result of verification using the device, it was confirmed that the liquid flowed into the air supply channel, and the damage to the device caused a pressure difference between the body cavity and the outside air, which flowed into the air supply channel.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|
| |
|
Search Alerts/Recalls
|