During observation of the flat filter, whitening was confirmed at the root of the lug on the female connector.Next, to check the function, we have connected an iso80369-6 inspection gauge and confirmed that the connection is possible without any abnormalities.Also, when water was injected with a syringe, no leakage was observed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|