BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 29420
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter was completely broken.A 150/10 renegade hi-flo was selected for use.During preparation, protection tube was irrigated, however, a complete break of the catheter was noted upon retrieving.The procedure was completed with another device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.The shipping hoop was hydrated, and the device was removed.The device was inspected for any damage or irregularities.The device showed damage in the form of a fracture located 16cm from the hub.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Event or Problem Description
It was reported that the catheter was completely broken.A 150/10 renegade hi-flo was selected for use.During preparation, protection tube was irrigated, however, a complete break of the catheter was noted upon retrieving.The procedure was completed with another device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE HI-FLO
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15404187
Report Number2124215-2022-34668
Device Sequence Number4947398
Product Code KRA
UDI-Device Identifier08714729280651
UDI-Public08714729280651
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K140329
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2024
Device Model Number29420
Device Catalogue Number29420
Device Lot Number0028642337
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-