| Model Number |
29420 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the catheter was completely broken.A 150/10 renegade hi-flo was selected for use.During preparation, protection tube was irrigated, however, a complete break of the catheter was noted upon retrieving.The procedure was completed with another device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for evaluation.The shipping hoop was hydrated, and the device was removed.The device was inspected for any damage or irregularities.The device showed damage in the form of a fracture located 16cm from the hub.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter was completely broken.A 150/10 renegade hi-flo was selected for use.During preparation, protection tube was irrigated, however, a complete break of the catheter was noted upon retrieving.The procedure was completed with another device.No patient complications were reported.
|
| |
|