Section d6a: date of implant incorrectly provided on previous report.Date of implant is not applicable for heartmate touch kit.Manufacturer's investigation conclusion: the reported event of the heartmate touch app not accepting an increase of speed was not confirmed.The heartmate touch tablet (serial number (b)(6) was returned and evaluated at abbott.The returned heartmate touch tablet was functionally tested with a test wireless adapter, system controller, power module, and mock circulatory loop and was found to function as intended during analysis.The pump speed was changed several times during testing without issue.The system operated for an extended period of time with no issues observed.The reported event could not be reproduced during testing.It was noted during testing that the tablet clock was approximately 188 days behind; this is able to be resolved by setting the clock in the tablet settings and does not indicate an issue with the system.The framework file downloaded from the returned heartmate touch tablet (labeled logheartmateframework_) contained relevant data from the event date of 23aug2022.It should be noted that the framework file captured these events on the timestamp of 16feb2022 due to the tablet clock being approximately 188 days behind.The heartmate touch app was launched on (b)(6) 2022 at 11:07:40.The adapter was connected in the following event at 11:08:05.The heartmate 3 system controller, serial number (b)(6), was connected at 11:08:10.The framework file then captured data consistent with the wireless adapter being disconnected from the power module from 3:11:29 to 3:12:02, and no additional events were captured on the event date.No atypical events were observed in the framework file.The root cause of the reported event could not be conclusively determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate touch kit, serial number (b)(6), was shipped from abbott on 04oct2021.Heartmate 3 instructions for use (ifu), rev.B, chapter 7 ¿alarms and troubleshooting¿ and heartmate touch communication system quick start guide, rev.B, under ¿troubleshooting guide¿ explain how to address the heartmate touch app freezing and how to set the tablet clock if it is incorrect.Heartmate 3 instructions for use (ifu), rev.B, chapter 4, entitled ¿heartmate touch communication system¿, and the heartmate touch communication system quick start guide, rev.B, under ¿how to use the heartmate touch communication system¿ explain the operation of the heartmate touch system, including how to use the heartmate touch app.Heartmate touch communication system quick start guide, rev.B, under "quick start" and heartmate 3 instructions for use (ifu), rev.B, chapter 4 ¿heartmate touch communication system¿ state "before launching the heartmate touch app, setup the tablet (in settings) for the specific region, time zone, the language and time format that the tablet is going to be used." no further information was provided.The manufacturer is closing the file on this event.
|