| Model Number |
1804225-12 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the heavily tortuous, heavily calcified distal left anterior descending artery that is 90% stenosed.The 2.25x12mm xience skypoint stent delivery system failure to cross due to anatomy.During removal resistance was met with anatomy.A non-abbott device used to complete the procedure.There were no adverse patient effects and no clinically significant delay reported.No additional information was provided.
|
| |
|