• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TOXO IGG; ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TOXO IGG; ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII Back to Search Results
Catalog Number 07028008190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with elecsys toxo igg on a cobas 8000 e 801 module.A questionable result from the sample was reported outside of the laboratory to a medic, who requested a rerun.The sample initially resulted in a toxo igg value of 11.5 iu/ml (reactive) on the e 801 analyzer.The sample was repeated on the e 801 analyzer on (b)(6) 2022, resulting in a value of 12.1 iu/ml (reactive).The sample was sent to two other laboratories for testing on the vidas and alinity platforms.It is not known specifically which platform was used at each laboratory.At the first laboratory, the toxo igg result was < 1 iu/ml (negative) on (b)(6) 2022.At the second laboratory, the toxo igg result was 0.40 iu/ml (negative) on (b)(6) 2022.The e 801 analyzer serial number is (b)(4).
 
Additional Manufacturer Narrative
The last calibration performed on (b)(6) 2022 was within expected ranges.The investigation could not identify a product problem.The cause of the event could not be determined.A general reagent issue can be excluded.The patient sample was requested for investigation, but the remaining volume was too low.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TOXO IGG
Common Device Name
ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15405902
Report Number1823260-2022-02807
Device Sequence Number12172329
Product Code LGD
UDI-Device Identifier04015630940301
UDI-Public04015630940301
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K073501
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07028008190
Device Lot Number598234
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-