| Model Number |
N/A |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A customer from outside the united states reported observation of falsely-depressed atellica im thyroid stimulating hormone 3-ultra (tsh3-ul) results which were discordant relative to alternate-method testing.The instructions for use (ifu) under the interpretation of results section states the following: "results of this assay should always be interpreted in conjunction with the patient's medical history, clinical presentation, and other findings." siemens healthcare diagnostics is investigating.
|
| |
|
Event or Problem Description
|
|
The customer reports observation of falsely-depressed atellica im thyroid stimulating hormone 3-ultra (tsh3-ul) results which were discordant relative to alternate-method testing.The customer has observed two instances where the atellica im tsh3-ul result is below the assay's measuring interval (<0.008 miu/l), but testing using an alternate method produced a higher result.The alternate-method results were considered consistent with the patient's clinical presentation, and were accepted as correct.There are no allegations of patient harm, and no reports of modification of treatment in association with the observed result discordance.
|
| |
|
Additional Manufacturer Narrative
|
|
Mdr 1219913-2022-00267 was initially filed on 2022-09-12.Additional information, 2022-10-13: siemens has concluded the investigation.A customer from outside the united states reported observation of falsely-depressed atellica im thyroid stimulating hormone 3-ultra (tsh3-ul) results which were discordant relative to alternate-method testing.Siemens reviewed the available information, and concluded that extremely rare genetic variants are the most likely cause of the undetectable tsh3-ul results.The instructions for use (ifu) for atellica im thyroid stimulating hormone 3-ultra (tsh3-ul) states the following, under limitations: "as with any immuno-recognition measurement of a peptide, extremely rare genetic variants may exhibit varying degrees of detection." a product performance issue has not been identified.The customer is operational; no further investigation is required.Note: in section h6, the codes for investigation findings and investigation conclusion have been updated.
|
| |
|
Search Alerts/Recalls
|