| Model Number |
1011-0139-05 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed for high impedance.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the internal paddles and they performed to specification.The internal paddles were put through extensive testing including resistance testing using a multimeter without duplicating the report.The paddles were returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|