SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67PFS50
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During inspection the they plugged in hose and power cord and power this device.No, the 36c was selected and running one hour on this temp, selecting and running on the others temp.With no error occurring.Unable to confirm customer reported problem.Unable to confirm customer reported problem.Replace the hose and the power cord.Replace the receptacle cable like a preventive maintenance.Changed the filter, performing a three hour test without any over temp to occur.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported during a surgery the equator reached a temperature of 60c (temperature shown on the screen display).As a result, the patient got burned on his left shoulder.The patient received regional anesthesia.It was the patient itself who shouted that he was getting burned.From this moment the doctors turned the machine off and removed it from the operation room.No medical or surgical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15406714
Report Number3012307300-2022-18797
Device Sequence Number14862892
Product Code JOH
UDI-Device Identifier10351688518644
UDI-Public10351688518644
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67PFS50
Device Catalogue Number67PFS50
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Initial Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-