| Model Number |
21-2766-24 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The device and a picture were received for evaluation.Visual and functional testing were performed.Visual inspection found the returned device was decontaminated with its original packaging.Functional testing confirmed the reported failure mode.The device was tested for occlusion by introducing water with a syringe; the device was occluded.The root cause of the reported issue was excess solvent in the component bonding operation.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the customer attempted to perform a priming of the product during the pre-use check, but could not do it successfully because he felt resistance through the tubing.No patient injury was reported.
|
| |
|