Model Number REXERIS-CF27
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.The device was returned to olympus for evaluation where service confirmed the customer's complaint.It was found that the ceramic was broken with missing fragments, and the outer tube was dented.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.Dhrs provided do not include packaging or labeling operations.For this reason, the final manufacture date is april 2016.Based on the results of the investigation, the damage which occurred was most likely caused by improper handling of the device during reprocessing, such as impact-dropped.A definitive root cause cannot be identified.This information is addressed in the instructions for use (ifu): "study this manual and other labeling thoroughly for safe handling, storage and usage, including instructions for all generators and accessories.Failure to properly follow the instructions, warnings, and cautions may lead to serious surgical consequences or injury to the patient.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments, or allow them to be struck by other objects." (warnings#1, page 4); " do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged.Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip.If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use." olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
The customer reported to olympus that the ceramic tip was damaged.There was no harm or user injury reported due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPAIR-EXCHANGE, ROTATING CF RESECTOSCOPE INNER SHEATH
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
9600 louisiana avenue north
brooklyn park, MN 55445
5082077661
MDR Report Key15406883
Report Number3011050570-2022-00112
Device Sequence Number6840004
Product Code HIH
UDI-Device Identifier00821925008298
UDI-Public00821925008298
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K890328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberREXERIS-CF27
Device Lot NumberRXDW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2021
Initial Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-