| Model Number |
862442 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the intellivue multi measurement server (mms) had a spo2 malfunction.The mms disconnected itself from the monitor.The device was in use at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A follow up report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
Problem statement: it was reported that the intellivue multi measurement server (mms) had a spo2 malfunction.The mms disconnected itself from the monitor.It is unknown if the product was in use on a patient, but no adverse event to the patient or user was reported.Evaluation: the product associated with the event was returned to bench repair for evaluation/analysis.The engineer determined that the spo2 was functioning as expected, but the measurement server (mms) disconnects itself when taking an nbp reading.An additional case was opened to address the non-invasive blood pressure (nbp) issue (see pr# (b)(4)).There was no trouble found with the spo2.Resolution and conclusion: the engineer determined that the spo2 met specifications and was functioning as expected.The customer has requested the device be returned.
|
| |
|