Catalog Number 400SMTXSFT0203
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2022
Event Type  malfunction  
Event Description
The patient was undergoing a coil embolization procedure in the middle cerebral artery (mca) using penumbra smart coils (smart coil) and a penumbra smart coil detachment handle (handle).During the procedure, the physician successfully detached smart coils in the target vessel using the handle.The physician then advanced another smart coil to the target vessel and attempted to detach it using the handle; however, the coil failed to detach after multiple attempts.Subsequently, while retracting the smart coil, the smart coil unintentionally detached.Therefore, the physician advanced the detached smart coil into the target location using the pusher assembly of the coil.The procedure was completed using non-penumbra coils.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15407040
MDR Text Key305191493
Report Number3005168196-2022-00424
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548015279
UDI-Public814548015279
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number400SMTXSFT0203
Device Lot NumberF112857
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received08/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial