Catalog Number LITNG7XTORQ130
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 08/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was undergoing a thrombectomy procedure in an artery of the right leg using an indigo system aspiration catheter 7 (cat7), lighting aspiration tubing (lightning), and a non-penumbra sheath.It was noted that the patient's anatomy was tortuous.During the procedure, the physician experienced resistance while advancing the cat7 through the sheath to the target location and subsequently, broke that cat7 at the mid-shaft within the sheath.The physician then used a snare device to remove the broken part of the cat7.It was also reported that there was perforation of the aorta that was then stented.It is unknown if the perforation was related to the cat7.The procedure was completed with an indigo system aspiration catheter 8 (cat8) and a new non-penumbra sheath.
 
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Additional Manufacturer Narrative
Perforation is included as a possible complication in the instructions for use (ifu) for the indigo aspiration system.Evaluation of the returned cat7 confirmed that the catheter was fractured.Based on the damage at the fracture location, this break was likely the result of torquing the device against resistance.If the cat7 is over-rotated against resistance, the device may become fractured.Further evaluation revealed ovalization at the distal tip of the catheter.This damage was incidental to the reported complaint and the root cause could not be determined.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM LIGHTNING 7
Common Device Name
QEW
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15407099
Report Number3005168196-2022-00423
Device Sequence Number6745111
Product Code QEW
UDI-Device Identifier00815948022584
UDI-Public815948022584
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K210083
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2025
Device Catalogue NumberLITNG7XTORQ130
Device Lot NumberH12905
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-