MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD CATHENA¿ SAFETY IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD CATHENA¿ SAFETY IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Model Number 386863
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the needle pierced through the bd cathena¿ safety iv catheter while introducing it.The following information was provided by the initial reporter: "the iv was started in the preoperative area, the stylette and catheter separated inappropriately.Anecdotally, the anesthesia assistant that witnessed the event reported that she has seen this before - yet it was unreported.".
 
Event or Problem Description
It was reported that the needle pierced through the bd cathena¿ safety iv catheter while introducing it.The following information was provided by the initial reporter: "the iv was started in the preoperative area, the stylette and catheter separated inappropriately.Anecdotally, the anesthesia assistant that witnessed the event reported that she has seen this before - yet it was unreported.".
 
Additional Manufacturer Narrative
H6: investigation summary: three photos and one used actual sample with open packaging were received by our quality team for evaluation.From the photos, the needle was observed to be pierced through the catheter at the nose of the catheter adapter.The sample was subjected to visual inspection.The needle was observed to be pierced through the catheter at the nose of the catheter adapter.A blood stain was observed in the catheter tubing, so the sample was used.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The manufacturing process was reviewed and investigated.No station was found which could potentially lead to the reported defect.As the returned sample was used, the needle piercing through the catheter most likely occurred during product application when the product was manipulated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CATHENA¿ SAFETY IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15407236
Report Number8041187-2022-00532
Device Sequence Number14858022
Product Code FOZ
UDI-Device Identifier00382903868636
UDI-Public00382903868636
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K201717
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number386863
Device Catalogue Number386863
Device Lot Number1294876
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer10/11/2022
Initial Report FDA Received Date09/12/2022
Supplement Report FDA Received Date10/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-