• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FINELINE II STEROX; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FINELINE II STEROX; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4459
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081); Extracardiac stimulation (1463); Defective Device (2588); High Capture Threshold (3266)
Health Effect - Clinical Code Device Overstimulation of Tissue (1991)
Date of Event 08/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited noise, high pacing threshold measurements, and intermittent loss of capture (loc).The noise was more pronounced in the bipolar configuration and less so in unipolar.The automatic capture feature mitigated the intermittent loc.Additionally, muscle/pocket stimulation was reported and an insulation issue was suspected on the rv lead.The pacemaker had reached normal end of life (eol) battery status, so during the device replacement procedure, this rv lead, and the chronic right atrial (ra) lead, were surgically abandoned and an entirely new pacemaker system was implanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINELINE II STEROX
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15407404
Report Number2124215-2022-35572
Device Sequence Number6840020
Product Code DTB
UDI-Device Identifier00802526061004
UDI-Public00802526061004
Combination Product (Y/N)Y
Initial Reporter CountryBE
PMA/510(K) Number
P960004/S014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/06/2007
Device Model Number4459
Device Catalogue Number4459
Device Lot Number434346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Initial Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;