| Model Number |
SSX-340 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Company representative reported capsular contracture on right side.Baker grade is unknown.It is unknown whether or not the device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Additional, changed, and/or corrected data: a2, a4, b5, d1, d2, d4, d6a, g4, h4, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iii.Device remain implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event that is not related to the device.Additional observations: crease fold is observed.White particles are observed.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade - iii".The device has been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|