• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC WEBSTER ® CS CATHETER WITH EZ STEER® THECHNOLOGY AND AUTO ID; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC WEBSTER ® CS CATHETER WITH EZ STEER® THECHNOLOGY AND AUTO ID; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE Back to Search Results
Model Number BD710FJ282CT
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 08/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Biosense webster manufacturer's reference number (b)(4) has two complaints that are related to the same incident.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a 69-year-old male patient (110 kgs) underwent a paroxysmal atrial fibrillation (afib)ablation procedure with a navistar® rmt thermocool® electrophysiology catheter and a webster ® cs catheter with ez steer® technology and auto id.The patient suffered cardiac tamponade requiring pericardiocentesis and prolonged hospitalization.It was reported that after the ablation, the patient suffered a pericardial effusion.It was discovered while the patient was still on the procedure table, after catheter removal, using transthoracic echo.The caller was uncertain if the patient's blood pressure was affected.A pericardiocentesis was ongoing at the time of the call.The patient remained stable.The physician did not indicate that they believed biosense webster, inc.Products contributed to the patient event.Additional information was received.This adverse event was discovered post use of biosense webster products.Pericardiocentesis was performed, no other intervention necessary.The patient¿s current condition is unknown, but patient went home.Patient stayed in hospital for 2 days, would have only been here for the day if procedure went according to plan.Transseptal puncture was performed.There was no known steam pop.The biosense webster representative is not sure when the event occurred, as it was noticed after the ablation was over.The biosense webster representative sent all catheters back because they were not sure which one would be under investigation at the time and wanted to cover all bases.Since sending them back, the physician has stated she suspects she perforated with the coronary sinus catheter at the beginning of the case, but they were never able to get a definitive diagnosis as the chest was never opened.The biosense webster representative does not know which coronary sinus catheter is the suspected device.They think both were used in an attempt to get the coronary sinus access but can not remember the order they were used in.Conservatively the bwi webster ® cs catheter with ez steer® technology and auto id and the a navistar® rmt thermocool® electrophysiology catheter were assessed as mdr reportable as ¿the physician has stated she suspects she perforated with the coronary sinus catheter at the beginning of the case, but they were never able to get a definitive diagnosis as the chest was never opened.¿ therefore, it is difficult to determine which device contributed to the event.
 
Additional Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).On 13-sep-2022 it was discovered that the device was received on 30-aug-2022 and the information was iadvertently omitted from the 3500a initial.As such the following fields have been processed: -h3.Reason for non- evaluation from device evaluation anticipated, but not yet begun to no -h3.Reason for non- evaluation from blank to other -h10.Additional manufacturer narrative should have included: the bwi product analysis lab received the device for evaluation on 30-aug-2022.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.In addition, during an internal review on 14-sep-2022, a correction was noted to the 3500a initial as we reported as a concomitant product ¿reprocessed 6fr cs, d, 10 pole¿; however, on 06-sep-2022, clarification had been received correcting this device to ¿reprocessed 6fr cs, f, 10 pole¿.Therefore, updated the ¿d10.Concomitant medical products and therapy dates¿ field.
 
Additional Manufacturer Narrative
The device evaluation was completed on 28-sep-2022.It was reported that a 69-year-old male patient (110 kgs) underwent a paroxysmal atrial fibrillation (afib)ablation procedure with a navistar® rmt thermocool® electrophysiology catheter and a webster ® cs catheter with ez steer® technology and auto id.The patient suffered cardiac tamponade requiring pericardiocentesis and prolonged hospitalization.The product was returned to biosense webster for evaluation.Bwi conducted a general inspection through the visual inspection and all features of the catheter tests.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the c3 ez steer cs with auto id catheter.Per the event, several tests were performed.The magnetic features were tested, and no issues were observed.In addition, the product was deflecting correctly.No malfunctions were observed during the product analysis.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.As part of bwi¿s quality process all devices are manufactured, inspected, and released to approved specifications.No malfunction was observed during the product analysis.The instructions for use states that careful catheter manipulation must be performed to avoid cardiac damage, perforation, or tamponade.The root cause of the adverse event remains unknown.There may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEBSTER ® CS CATHETER WITH EZ STEER® THECHNOLOGY AND AUTO ID
Common Device Name
ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez
MX  
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15408207
Report Number2029046-2022-02208
Device Sequence Number20571115
Product Code DRF
UDI-Device Identifier10846835002423
UDI-Public10846835002423
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K090898
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/24/2023
Device Model NumberBD710FJ282CT
Device Catalogue NumberBD710FJ282CT
Device Lot Number30342546M
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer09/13/2022
09/28/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date09/29/2022
10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NAVISTAR RMT THERMOCOOL; PENTARAY NAV ECO 7FR, D, 2-6-2; REPROCESSED 6FR CS, D, 10 POLE; REPROCESSED 6FR CS, F, 10 POLE; UNK_SMARTABLATE GENERATOR
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age69 YR
Patient SexMale
Patient Weight110 KG
-
-