| Model Number |
02.12.0022L |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 22-aug-2022: lot 2101922: (b)(4) items manufactured and released on 25-mar-2021.Expiration date: 2026-03-14.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Additional components involved: gmk-sphere 02.07.1202l tibial tray fixed cemented size 2 l (k090988) lot 2008493: (b)(4) items manufactured and released on 03-dec-2020.Expiration date: 2025-11-24.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Gmk-sphere 02.12.E001rp patella resurfacing size 1 e-cross (k202022) lot 2002542: (b)(4) items manufactured and released on 03-jul-2020.Expiration date: 2025-06-09.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Gmk-sphere 02.12.E0210fl tibial insert fixed sphere flex size 2/10 mm l e-cross (k202022) lot 2012370: (b)(4) items manufactured and released on 13-jan-2021.Expiration date: 2025-12-16.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.
|
| |
|
Event or Problem Description
|
|
At about 9 months after the primary surgery, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout, removed all implants, and implanted an antibiotic spacer.The surgery was completed successfully.
|
| |
|
Search Alerts/Recalls
|