| Model Number |
SONATA |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
It was reported that in mid (b)(6) 2021 the user had suddenly stopped hearing with the device.The user was re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage which is expected to have been present whilst implanted.The reported issue was most likely caused by the presence of an air pocket over the reference electrode.Mechanical damages found during investigation are attributable to the removal surgery.This is a final report.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2021.The user had suddenly stopped hearing with the device.The user was re-implanted.
|
| |
|