C.R. BARD, INC. (BASD) -3006260740 TITANIUM IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 9.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 0602230
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260); Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 07/2026).
 
Event or Problem Description
It was reported that during a port placement procedure, the device allegedly had suction issue.It was further reported that few solution that the syringe has comes out through the connection.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one bardport ti implantable port, one flushing connector loaded to a catheter segment, one introducer needle, two cath-locks, one 10.0fr peel-apart sheath and vessel dilator, one vein pick, one right-angle non-coring needle, one straight-angle non-coring needle, one guidewire straightener, one j-tip guidewire loaded in a guidewire hoop, one tunneler, one catheter in three segments and one safety infusion set were returned for evaluation.Gross, microscopic visual and functional testing were performed.The investigation is confirmed for the reported fluid leak, suction issue and identified fracture issues as multiple cracks were observed on the introducer needle hub.Upon infusion, a leak from the hub was observed; water exiting through the distal tip without issue.Furthermore, during the sample evaluation the introducer needle was noted to be bent was considered to be an incidental findings since it was not related to the original reported event.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.
 
Event or Problem Description
It was reported that during a port placement procedure, the device allegedly had suction issue.It was further reported that few solution that the syringe has comes out through the connection.The procedure was completed by using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITANIUM IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 9.6F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15409117
Report Number3006260740-2022-03568
Device Sequence Number6840077
Product Code LJT
UDI-Device Identifier00801741025563
UDI-Public(01)00801741025563
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K050310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0602230
Device Catalogue Number0602230
Device Lot NumberREGQ2960
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-