| Model Number |
8401 |
| Medical Device Problem Code |
Gas Output Problem (1266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The dhr review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.The device was received for evaluation.Visual and functional testing were performed.Visual inspection found tamper seal was intact; however, there was a cracked battery compartment and loose speaker.Functional testing was able to duplicate the reported problem.The flow rate read 55 ml/min but the monitor was still occluded and passed leak testing.The root cause of the reported issue was found to be a leak in the pneumatics through cracked filter inlet port.The back case was replaced.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.
|
| |
|
Event or Problem Description
|
|
It was reported that there was high co2, unstable.No patient injury was reported.
|
| |
|
Search Alerts/Recalls
|