| Model Number |
8401 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that co2 was reading high on the display.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Operator of device is unknown.No information has been provided to date.Initial reporter also sent report to fda is unknown.No information has been provided to date.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The reported complaint was not verified.The returned 8401 monitor was tested with a mass flow meter and was within specification of 74 mmhg (spec: 74+/-3 mmhg).Unit was inspected internally and no issues observed.No problem was detected.Monitor working as intended.This remediation mdr was generated under protocol (b)(4), as a result of warning letter (b)(4).
|
| |
|