| Model Number |
321-35-46 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that while reaming the glenoid, it fractured off the square piece that attaches to the tri drive.Rep believes the tech did not fully seat the reamer on the tri drive and when he started reaming.Nothing was damaged on the patients end.No fragments, parts or pieces fell into the patient.It fractured outside the 81 y/o male patient and from the tri drive connection piece most distal.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Device is returning.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the broken cannulated reamer reported may have been due to incomplete seating of the reamer on the tri-driver prior to use, which led to fracturing of the reamer post during reaming.Section h11: the following sections have corrected information: (h6) component code: 912, post.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury, nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|