| Model Number |
PV2015L20-A |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information about the event has been requested.According to provided information the device is available, but has not yet been returned to manufacturer.The customer has not been able to clearly identify the concerned batch and provided 3 potentially applicable batch numbers: 694239, 694241 or 684727.The catheter, which could be successfully inserted, had the same batch number as one of the 3 provided batch numbers.Current status of investigation indicates that picco catheter batch 694239 has been shipped to this hospital.Data in expiration date and device manufacture date has been added based on this status.A supplemental emdr will be sent when the investigation is completed.
|
| |
|
Event or Problem Description
|
|
It has been reported that during insertion via the femoral route on an already existing arterial catheter ("mandraining" on an already existing catheter"), the catheter for picco cracks when passing through the guide.The doctor was obliged to remove the catheter and guide, and to change the picco catheter.Success of the procedure by changing the catheter.No clinical consequences have been observed.Manufacturer reference: #(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Further information about the event has been requested.The customer is currently providing contradicting information if the guidewire or the catheter is defective.Further information incl.Pictures or videos have been requested, but not received.According to provided information the device is available, but has not yet been returned to manufacturer.A supplemental emdr will be sent when the investigation is completed.
|
| |
|
Event or Problem Description
|
|
Initial information indicated a broken catheter.New information received from the health facility indicated that the guidewire broke instead of the catheter."it seems to be the guide wire that has cracked.".No further information provided so far.No clinical consequences have been observed.Manufacturer reference:(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The complained catheter and guidewire have been returned and were inspected visually.The guidewire was still inserted in the catheter.The justification for reporting could not be confirmed for this case: visual inspection did not show any cracks of the guidewire like uncoiled or torn off.There were only small kinks on both guidewire tips.Also, no crack and leakage could be reproduced for the catheter.The catheter tubing only showed some minor dents next to the channel separation.Kinks on the guidewire and dents on the catheter tubing are indications for the use of force.Moreover, the malfunction has been detected when the catheter was inserted with the guidewire and not before insertion ("the catheter for picco cracks when passing through the guide").This supports the fact, that the catheter and guidewire did not show any deficiency before use.The dhr of all provided lots have been reviewed but did not reveal any non-conformities or deviations relevant to the reported issue.No further complaints have been received for these lots.There is no indication for a systematic root cause as a deficiency of design, production or material considering the very low complaint rate (< 0,01 %, considering all types of picco catheters).Based on the complaint investigation, it cannot be excluded that an handling error during use occurred.No further information regarding the circumstances during insertion of the catheter have been provided to further evaluate potential handling errors.The instruction for use (ifu) states warnings for the catheter placement procedure: "warning: kinks, bending or handling with a clamp can irreversibly damage the catheter." additionally, the ifu describes the insertion procedure seldinger technique.Overall, investigations did no indicate that the device failed to meet its specification when the event occurred.There is also no indication for a weakness of the supporting instruction for use.Upon the event occurrence the device was involved and catheter insertion procedure on a patient for advanced hemodynamic monitoring was ongoing.The accessory guidewire is supplied in the product picco catheter.The procedure to insert a catheter (¿seldinger technique¿) is state of the art for the healthcare professionals.The issue is monitored on a regularly basis in order to detect early trends.As there is no trend for this kind of issue no additional actions will be taken.With the information stated above the complaint will be closed.
|
| |
|
Event or Problem Description
|
|
Manufacturer reference: # (b)(4).
|
| |
|