SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 750160
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Sample was received with its original packaging.Cuff symmetry test was performed to the unit.When the cuff was inflated with 15cc of air, the cuff inflated completely around the tube, however, the cuff sees asymmetrical.Then when the cuff was measured and the percentage for the symmetry was 35.48 percent.This result is over 33.3 percent that is the minimum acceptable for air cuff.Based on the test, the unit is within spec.The complaint is not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.D5 operator of device is unknown this remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event or Problem Description
It was reported that during a pre-use test, the customer noticed the cuff did not inflate normally.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15410577
Report Number3012307300-2022-18925
Device Sequence Number6744710
Product Code JOH
UDI-Device Identifier15021312006407
UDI-Public15021312006407
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/11/2024
Device Model Number750160
Device Catalogue Number750160
Device Lot Number3904496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-