SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 750180
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during pre-test, part of the cuff was sticking to the cannula.No patient injury was reported.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.One picture was received for evaluation.Visual inspection of the picture found the cuff is inflated, and one part is sticking to the shaft.Although, no samples were received to perform inspection or testing, a probable root cause is teflon was not properly distributed on the cuff.The following actions were taken: an alert was generated to show the operators the procedure of teflon application in the cuff to prevent sticking to the shaft.Production personnel was trained on quality alert by an area trainer.A visual aid was released to show the operators the procedure of teflon application in the cuff to prevent sticking to the shaft.Production personnel was trained on quality alert by an area trainer.Operator of device is unknown.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15410612
Report Number3012307300-2022-18930
Device Sequence Number6999406
Product Code JOH
UDI-Device Identifier15021312006421
UDI-Public15021312006421
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number750180
Device Catalogue Number750180
Device Lot Number3760765
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-